IEC60898-1 MCCB dielectric properties Modular Intelligent Safety Tester
Standards & Clauses:
This tester conforms to the requirements of numerous standards, including but not limited to IEC 61009-1:2012 and IEC 60898-1:2015 Clause 9.7.
Featuring seven comprehensive functions—AC withstand voltage, DC withstand voltage, insulation resistance, grounding, leakage current, power, and low-voltage start testing—this advanced platform provides all-in-one safety testing capabilities.
With its globally advanced modular architecture, the system integrates safety compliance testing with intelligent information management. Supporting data collection, storage, analysis, and transmission, it offers a cutting-edge solution for smart, connected manufacturing environments and is the optimal choice for intelligent factory testing.
Technical Parameters:
| Model | Test Project | Function | Accuracy | ||||||||
| ACW | DCW | IR | GB | LC | PWR | STR | O/S | Variable frequency power supply | transformer | 1% | |
| SN6872F | √ | √ | √ | √ | √ | √ | √ | √ | √ | ||
| Rated output | 5KV/100mA, short-circuit current>200mA |
| The output voltage range | 100 ~ 5000V |
| Output voltage accuracy | ±(1%× set value +2 words) |
| Voltage resolution | 1V |
| Output frequency | 50/60Hz |
| Frequency accuracy | ±0.1Hz |
| Output adjustment | ±(1%× set value +2 words) |
| Current upper limit | 0.00 to 100.00mA |
| The current lower limit | 0.000 to 9.999mA |
| Current measurement range | 0.00 ~ 100.0mA |
| The current measurement resolution | 0.001/0.01mA |
| Current measurement accuracy | ±(1%× reading value +5 words) |
| Slow rise time | 0-999.9s |
| Slow down time | 0.1-999.9s |
| Test time | 0,0.3-999.9s |
| Time resolution | 0.1s |
| Time accuracy | ±(0.2% x set value +1 word) |
| Arc detection | 0-9 |
| Current compensation range | 0.000-20.00mA |
| Rated output | 6KV/10mA |
| The output voltage range | 200~6000V |
| Output voltage accuracy | ±(1%× set value +2 words) |
| The output ripple | <2% |
| Current upper limit | 0.00 to 10.00mA |
| The current lower limit | 0.000 to 9.999mA |
| Current measurement range | 0.00 ~ 10.0mA |
| The current measurement resolution | 0.001/0.01mA |
| Current measurement accuracy | ±(1%× reading value +5 words) |
| Slow rise time | 0-999.9s |
| Slow down time | 0.1-999.9s |
| Test time | 0,0.3-999.9s |
| Time resolution | 0.1s |
| Time accuracy | ±(0.2% x set value +1 word) |
| Arc detection | 0-9 |
| Current compensation range | 0.000-5.000mA |
| Discharge time | 100ms |
| Rated output | 2500V/50GΩ | |
| The output voltage range | 200~2500V | |
| Voltage resolution | 1V | |
| Output voltage accuracy | ±(1%× set value +2 words) | |
| Ripple coefficient | ≤2% | |
| Resistance measuring range | 0.00 to 10.00mA | |
| Resolution of resistance measurement | 0.000 to 9.999mA | |
| Accuracy of measurement |
1.000MΩ~999.9MΩ Output 200-500V |
±(7%× reading value +3 words) |
|
1.000MΩ~999.9MΩ Output 501-2500V |
±(5%× reading value +3 words) | |
|
1.00GΩ~10.00GΩ Output 501-2500V |
±(7%× reading value +3 words) | |
|
10.00GΩ~50.00GΩ Output 1000-2500V |
±(15%× reading value +3 words) | |
| Slow rise time | 0-999.9s | |
| Slow down time | 0.1-999.9s | |
| Test time | 0,0.3-999.9s | |
| Time resolution | 0.1s | |
| Time accuracy | ±(0.2% x set value +1 word) | |
| Discharge time | <100ms | |
| Rated output | 32A/600mΩ | |
| The output voltage range | 1.00~32.00A | |
| Voltage resolution | 0.01A | |
| Output voltage accuracy | ±(1%× set value +3words) | |
| Current fluctuations | ≤0.4% x Set value/min | |
| The output voltage | 3.00-8.00V AC | |
| The resolution of the | 0.1V | |
| The output frequency | 50Hz/60Hz | |
| Frequency accuracy | ±0.1Hz | |
| Upper/lower resistance setting | 0.1-600mΩ | |
| Resistance range | 0.1-600mΩ | 1.00~10.00A |
|
0.1 the Rm Ω R=(6400/ set current value)m ω |
10.10~64.00A | |
| Resolution of resistance measurement | 0.1/1mΩ | |
| Accuracy of resistance measurement | ±(1%× reading value +3 words) | |
| Voltage measuring range | 0.00-6.40V AC | |
| Voltage measurement resolution | 0.01V | |
| Accuracy of voltage measurement | ±(1%× reading value +3 words) | |
| Resistance compensation | 0-100mΩ | |
| The test of time | 0,0.3-999.9s | |
| Temporal resolution | 0.1s | |
| Time accuracy | ±(0.2%× reading value +3 words) | |
| Test Probe Setup | Ground contact current: G-L, G-N, AUTO (G-L, G-N) | |
| Contact current to surface: PH-L, PH-PL | ||
| 7 kinds of human body simulation circuits (MD) are standard (optional customized human body network) | MD A. |
GB12113/IEC60990,GB4706.1/IEC60335-1 GB7000.1/IEC60598-1,GB4943.1/IEC60950-1 GB8898.1/IEC60065 |
| MD B. | UL1563 | |
| MD C. | GB4943.1/IEC60950-1 | |
| MD D. |
UL544P
|
|
| MD E. | UL544NP/UL484/UL923/ UL1029/UL471/UL867/UL697 | |
| MD F. |
GB12113/IEC60990 GB7000.1/IEC60598-1 |
|
| MD G. | GB4943.1/IEC60950-1 | |
|
Test power supply
|
Polarity Switching: On, Off, Auto | |
| Grounded and Un-grounded Conversion: On, Off | ||
| The transition between the normal connection state and the non-connection state of the neutral line of the test power supply: ON, OFF | ||
| Test voltage output | Single-phase, requires an external isolation transformer or variable frequency power supply; provides the voltage and frequency required for the test. | |
| Test voltage upper limit setting | 0.0V~300.0V AC | |
| Test voltage lower limit setting |
0.0V~300.0V AC
|
|
| Test voltage resolution |
0.1V ; 45Hz~65Hz |
|
| Test voltage measurement range | 0.0V~300.0V,45Hz~65Hz | |
| Test voltage resolution | 0.1V,45Hz~65Hz | |
| Test voltage measurement accuracy | 20.0V~300.0V: ±(2%×reading value+3 digits) | |
| load current | 20A | |
| Overcurrent Protection | RMS current and response time: 25A<3s | |
| Peak current and response time: 70A<10ms | ||
| Touch current measurement upper limit (effective value) | 0.0mA~20.000mA | |
| Touch current measurement lower limit (RMS) | 0.0mA~20.000mA | |
| Touch current measurement resolution (RMS) | 0.001mA | |
| Touch current measurement judgment error (RMS) | AC | ±(1%×reading value+3 digits) |
|
15Hz≤f≤ 100kHz |
±(1%×reading value+3 digits) | |
|
100kHz<f≤ 1000kHz
|
±(2%×reading value+3 digits) | |
| Touch current compensation range | 0.000~1.000mA; automatic measurement; can be turned on or off | |
| testing time | 0; (1 ~999.9); 0 is infinite | |
| Test time resolution | 0.1s | |
| test accuracy | ±( 1% × set value + 1 word) | |
| test method | AUTO (G-L, G-N); time for 1 and a half times each | |
| Power upper/lower limit settings | 0.0~6000W | |
| Power measurement range | 0.0-999.9W,1000-6000W | |
| Power measurement resolution | 0.1W,1W | |
| Power Measurement Accuracy | PF>0.9 | ±(0.1%×reading value+0.1%×range) |
| PF≤0.9 | ±(0.4%×reading value+0.1%×range) | |
| Voltage measurement range | 0.0-300.0V | |
| Voltage measurement resolution | 0.1V | |
| Voltage measurement accuracy | ±(0.1%×reading value+0.1%×range) | |
| testing time | 0,0.3~999.9s | |
| Time resolution | 0.1s | |
| time accuracy | ±(0.2%×setting value+1 word) | |
| Current upper/lower limit setting | 0.00-20.00A |
| Current measurement range | 0.000-9.999A ,10.00-20.00A |
| Current measurement resolution | 0.001A/0.01A |
| Current measurement accuracy | ±(0.1%×reading value+0.1%×range) |
| Voltage measurement range | 0.0-300.0V |
| Voltage measurement resolution | 0.1V |
| Voltage measurement accuracy | ±(0.1%×reading value+0.1%×range) |
| Testing time | 0,0.3~999.9s |
| Time resolution | 0.1s |
| time accuracy | ±(0.2%×setting value+1 word) |
|
|
| MOQ: | 1 |
| Price: | Customized |
| Standard Packaging: | Plywood |
| Delivery Period: | 30 Days |
| Payment Method: | L/C, , T/T, Western Union, MoneyGram |
| Supply Capacity: | 1 set per month |
IEC60898-1 MCCB dielectric properties Modular Intelligent Safety Tester
Standards & Clauses:
This tester conforms to the requirements of numerous standards, including but not limited to IEC 61009-1:2012 and IEC 60898-1:2015 Clause 9.7.
Featuring seven comprehensive functions—AC withstand voltage, DC withstand voltage, insulation resistance, grounding, leakage current, power, and low-voltage start testing—this advanced platform provides all-in-one safety testing capabilities.
With its globally advanced modular architecture, the system integrates safety compliance testing with intelligent information management. Supporting data collection, storage, analysis, and transmission, it offers a cutting-edge solution for smart, connected manufacturing environments and is the optimal choice for intelligent factory testing.
Technical Parameters:
| Model | Test Project | Function | Accuracy | ||||||||
| ACW | DCW | IR | GB | LC | PWR | STR | O/S | Variable frequency power supply | transformer | 1% | |
| SN6872F | √ | √ | √ | √ | √ | √ | √ | √ | √ | ||
| Rated output | 5KV/100mA, short-circuit current>200mA |
| The output voltage range | 100 ~ 5000V |
| Output voltage accuracy | ±(1%× set value +2 words) |
| Voltage resolution | 1V |
| Output frequency | 50/60Hz |
| Frequency accuracy | ±0.1Hz |
| Output adjustment | ±(1%× set value +2 words) |
| Current upper limit | 0.00 to 100.00mA |
| The current lower limit | 0.000 to 9.999mA |
| Current measurement range | 0.00 ~ 100.0mA |
| The current measurement resolution | 0.001/0.01mA |
| Current measurement accuracy | ±(1%× reading value +5 words) |
| Slow rise time | 0-999.9s |
| Slow down time | 0.1-999.9s |
| Test time | 0,0.3-999.9s |
| Time resolution | 0.1s |
| Time accuracy | ±(0.2% x set value +1 word) |
| Arc detection | 0-9 |
| Current compensation range | 0.000-20.00mA |
| Rated output | 6KV/10mA |
| The output voltage range | 200~6000V |
| Output voltage accuracy | ±(1%× set value +2 words) |
| The output ripple | <2% |
| Current upper limit | 0.00 to 10.00mA |
| The current lower limit | 0.000 to 9.999mA |
| Current measurement range | 0.00 ~ 10.0mA |
| The current measurement resolution | 0.001/0.01mA |
| Current measurement accuracy | ±(1%× reading value +5 words) |
| Slow rise time | 0-999.9s |
| Slow down time | 0.1-999.9s |
| Test time | 0,0.3-999.9s |
| Time resolution | 0.1s |
| Time accuracy | ±(0.2% x set value +1 word) |
| Arc detection | 0-9 |
| Current compensation range | 0.000-5.000mA |
| Discharge time | 100ms |
| Rated output | 2500V/50GΩ | |
| The output voltage range | 200~2500V | |
| Voltage resolution | 1V | |
| Output voltage accuracy | ±(1%× set value +2 words) | |
| Ripple coefficient | ≤2% | |
| Resistance measuring range | 0.00 to 10.00mA | |
| Resolution of resistance measurement | 0.000 to 9.999mA | |
| Accuracy of measurement |
1.000MΩ~999.9MΩ Output 200-500V |
±(7%× reading value +3 words) |
|
1.000MΩ~999.9MΩ Output 501-2500V |
±(5%× reading value +3 words) | |
|
1.00GΩ~10.00GΩ Output 501-2500V |
±(7%× reading value +3 words) | |
|
10.00GΩ~50.00GΩ Output 1000-2500V |
±(15%× reading value +3 words) | |
| Slow rise time | 0-999.9s | |
| Slow down time | 0.1-999.9s | |
| Test time | 0,0.3-999.9s | |
| Time resolution | 0.1s | |
| Time accuracy | ±(0.2% x set value +1 word) | |
| Discharge time | <100ms | |
| Rated output | 32A/600mΩ | |
| The output voltage range | 1.00~32.00A | |
| Voltage resolution | 0.01A | |
| Output voltage accuracy | ±(1%× set value +3words) | |
| Current fluctuations | ≤0.4% x Set value/min | |
| The output voltage | 3.00-8.00V AC | |
| The resolution of the | 0.1V | |
| The output frequency | 50Hz/60Hz | |
| Frequency accuracy | ±0.1Hz | |
| Upper/lower resistance setting | 0.1-600mΩ | |
| Resistance range | 0.1-600mΩ | 1.00~10.00A |
|
0.1 the Rm Ω R=(6400/ set current value)m ω |
10.10~64.00A | |
| Resolution of resistance measurement | 0.1/1mΩ | |
| Accuracy of resistance measurement | ±(1%× reading value +3 words) | |
| Voltage measuring range | 0.00-6.40V AC | |
| Voltage measurement resolution | 0.01V | |
| Accuracy of voltage measurement | ±(1%× reading value +3 words) | |
| Resistance compensation | 0-100mΩ | |
| The test of time | 0,0.3-999.9s | |
| Temporal resolution | 0.1s | |
| Time accuracy | ±(0.2%× reading value +3 words) | |
| Test Probe Setup | Ground contact current: G-L, G-N, AUTO (G-L, G-N) | |
| Contact current to surface: PH-L, PH-PL | ||
| 7 kinds of human body simulation circuits (MD) are standard (optional customized human body network) | MD A. |
GB12113/IEC60990,GB4706.1/IEC60335-1 GB7000.1/IEC60598-1,GB4943.1/IEC60950-1 GB8898.1/IEC60065 |
| MD B. | UL1563 | |
| MD C. | GB4943.1/IEC60950-1 | |
| MD D. |
UL544P
|
|
| MD E. | UL544NP/UL484/UL923/ UL1029/UL471/UL867/UL697 | |
| MD F. |
GB12113/IEC60990 GB7000.1/IEC60598-1 |
|
| MD G. | GB4943.1/IEC60950-1 | |
|
Test power supply
|
Polarity Switching: On, Off, Auto | |
| Grounded and Un-grounded Conversion: On, Off | ||
| The transition between the normal connection state and the non-connection state of the neutral line of the test power supply: ON, OFF | ||
| Test voltage output | Single-phase, requires an external isolation transformer or variable frequency power supply; provides the voltage and frequency required for the test. | |
| Test voltage upper limit setting | 0.0V~300.0V AC | |
| Test voltage lower limit setting |
0.0V~300.0V AC
|
|
| Test voltage resolution |
0.1V ; 45Hz~65Hz |
|
| Test voltage measurement range | 0.0V~300.0V,45Hz~65Hz | |
| Test voltage resolution | 0.1V,45Hz~65Hz | |
| Test voltage measurement accuracy | 20.0V~300.0V: ±(2%×reading value+3 digits) | |
| load current | 20A | |
| Overcurrent Protection | RMS current and response time: 25A<3s | |
| Peak current and response time: 70A<10ms | ||
| Touch current measurement upper limit (effective value) | 0.0mA~20.000mA | |
| Touch current measurement lower limit (RMS) | 0.0mA~20.000mA | |
| Touch current measurement resolution (RMS) | 0.001mA | |
| Touch current measurement judgment error (RMS) | AC | ±(1%×reading value+3 digits) |
|
15Hz≤f≤ 100kHz |
±(1%×reading value+3 digits) | |
|
100kHz<f≤ 1000kHz
|
±(2%×reading value+3 digits) | |
| Touch current compensation range | 0.000~1.000mA; automatic measurement; can be turned on or off | |
| testing time | 0; (1 ~999.9); 0 is infinite | |
| Test time resolution | 0.1s | |
| test accuracy | ±( 1% × set value + 1 word) | |
| test method | AUTO (G-L, G-N); time for 1 and a half times each | |
| Power upper/lower limit settings | 0.0~6000W | |
| Power measurement range | 0.0-999.9W,1000-6000W | |
| Power measurement resolution | 0.1W,1W | |
| Power Measurement Accuracy | PF>0.9 | ±(0.1%×reading value+0.1%×range) |
| PF≤0.9 | ±(0.4%×reading value+0.1%×range) | |
| Voltage measurement range | 0.0-300.0V | |
| Voltage measurement resolution | 0.1V | |
| Voltage measurement accuracy | ±(0.1%×reading value+0.1%×range) | |
| testing time | 0,0.3~999.9s | |
| Time resolution | 0.1s | |
| time accuracy | ±(0.2%×setting value+1 word) | |
| Current upper/lower limit setting | 0.00-20.00A |
| Current measurement range | 0.000-9.999A ,10.00-20.00A |
| Current measurement resolution | 0.001A/0.01A |
| Current measurement accuracy | ±(0.1%×reading value+0.1%×range) |
| Voltage measurement range | 0.0-300.0V |
| Voltage measurement resolution | 0.1V |
| Voltage measurement accuracy | ±(0.1%×reading value+0.1%×range) |
| Testing time | 0,0.3~999.9s |
| Time resolution | 0.1s |
| time accuracy | ±(0.2%×setting value+1 word) |